Portable Respirator
Microvent CPR
A device that assists or controls the patient's breathing by supplying a certain amount of oxygen
Chosun Instrument Inc.
o
As a time/volume type
ventilator using compressed oxygen as a power source, both automatic and manual
types are available.
o
Automatic/manual mode
changeover switch is attached separately, so immediate mode change is possible.
o
In the case of the automatic
method, oxygen is supplied repeatedly by adjusting the number of breaths per
minute (BPM) and the amount of oxygen supplied per cycle (Tidal Volume).
o
The number of breaths per
minute is 10 times, and the amount of oxygen supplied per minute can be
adjusted from 150ml to 600ml. (Oxygen supply amount and frequency can be finely
adjusted)
o
The intake/exhaust ratio (I :
E) of the respiratory system is maintained at 1:5±20%.
o
In the case of the manual
method, oxygen is supplied only when the trigger type oxygen supply handle
(Manual Trigger) is operated, so it can be used in parallel with CPR. (Supply
of 40ℓ of oxygen per minute)
o
It is in the form of supplying
oxygen by simply pulling the trigger-shaped oxygen supply handle, so oxygen can
be supplied more conveniently and conveniently.
o
When the airway pressure rises
above 45cmH2O, such as when the patient's airway is blocked, the alarm function
operates to discharge the excess pressure outside and informs the user by
sounding an alarm.
o
Even if the pressure in the
airway increases, the excess pressure (over 45cmH2O) is discharged to the
outside, but oxygen must be supplied during the specified inspiration time.
o
For more accurate oxygen supply
to the patient, you can select the setting range of flow rate and respiratory
rate more precisely.
o
In order to prevent leakage of
oxygen gas that may occur between the mask and the patient's face, it is
possible to supply oxygen to the patient using the oxygen supply handle while
maintaining the airtightness of the mask while supporting the chin with both
hands.
o
A One-Way valve is installed to
prevent contamination of the respiratory tract by vomit, and cleaning and
replacement are possible when contaminated.
o
Operating temperature and
humidity : -18℃∼50℃, 40%∼95%
o
Storage temperature and
humidity: -40℃∼60℃, 40%∼95%
o
Mask connector inner/outer
diameter: 15mm/22mm
o
MicroVenT Handset size: 120×55×100mm
o
MicroVenT Handset Weight: 275g
o
Size of carrying bag:
560×220×210mm ± 10%
o
Total product weight: 8.0kg ±
10%